AM8060 twin channel thermometer – displays both probes simultaneously plus Channel 1 minus Channel 2. This unit is ideal to perform calibration bath surveys and when testing more than one item at a time. It’s also ideal as you can characterise 1 channel with a top of the range SPRT probe and the other channel with a working Secondary SPRT probe.
The AM8060 works with 100 ohm or 25 ohm probes or a combination of the two.
The AM8040 is a cheaper option than the AM8060 as its single channel.
Both the above thermometers are ideal for a secondary lab – a true 3 decimal resolution. Easy to use and affordable. Connection is via gold plated copper spade plugs which require careful use but do provide less resistance.
Resistance thermometers are classified as
AccuMac make some of the best probes in the world. They continue to provide probes to Primary labs worldwide – e.g. PTB (Germany National Metrology Lab), MSL of New Zealand and the Nepal National Metrology Lab.
Many Primary labs would use a thermometry bridge (6-10 decimal places) with their probes rather than an instrument like the AM8060. The AM8060 is fabulous for a secondary lab.
AccuMac make both 100 and 25 ohm probes in various diameters, lengths and sheaths.
If what Teltherm quotes doesn’t suit let us know and we can get you prices for other options.
Some info re temperature probes that may help you with your choice.
ITS-90 requires that the SPRT/PRT has W(Ga)>=1.11807?
All the PRT readouts read a PRT’s resistance value and converts that value to temperature based on the ITS-90 formula or CVD formula etc.
We say 100 ohm SPRTs are more sensitive, which means they have a better resolution for the same amount of temperature change.
For example, for roughly every 1 °C change in temperature, the 100 ohm SPRT will have a 0.4 ohm change in resistance, while a 25 ohm SPRT will only have a 0.1 ohm change in resistance.
So the resistance value changes 4 times larger from a 100 ohm SPRT than that from a 25 ohm SPRT, thus the PRT can produce more refined readings (better resolution), hence more sensitive.
This is not a factor in many calibration labs as the readouts such as Thermometry Bridge etc. can read resistance up to 6-10 decimal places. But it can be an issue for commercial users or field users where the readouts are not as advanced - you often see primary labs use 25 ohm SPRTs but you rarely see anything below 100 ohm from an industrial user.
One clear advantage of 25 ohm SPRT is stability, which means they are more durable and stable over time. This is because 25 ohm SPRTs use thicker/bigger platinum wires to construct the sensing element than compared to 100 ohm SPRTs.
Thicker/bigger wires are more shock resistant, and hence more stable.
Because AccuMac are a PRT manufacturer, they know the intricacies to balance performance and stability.
We recommend the following probes
AM1880 bird cage design (inconel sheath)
AM1900 (quartz sheath)
AM1800 series (inconel sheath)
AM1968 Platinum Capsule SPRT for cryogenic work (-260°C to 232 °C).
We recommend the following thermometers:
AM8060
We recommend the following probes:
AM1850-25 – use as a reference if you have the budget
AM1760 series – great working SPRT probes
AM1750 series – affordable secondary reference PRTS
Once you have purchased your new AccuMac equipment please ask your calibration Laboratory to do the following system calibration to achieve the best accuracy from you AccuMac thermometer and probe:
If you have purchased a top quality SPRT probe, the Primary lab is likely to anneal the probe first, AccuMac suggests the following:
The Primary Lab may perform a stabilization test of the SPRT first.
Normally this would be a 2 hour anneal at 660-670 °C.
This is very necessary because no matter how stable the SPRT is at AccuMac’s lab, a probe will get ‘stressed’ after long distance transportation. A 2-hour anneal will release this stress.
Please inform your Primary lab the annealing temperature needs to be 660-670 °C.
Some labs mistakenly anneal their SPRTs at 420 °C, which is not correct. Annealing at least to 500 °C is preferable.
Measurement Standards Laboratory in New Zealand and Teltherm Laboratory are very familiar with this process and currently charge from $1100 per probe to calibrate new ‘AccuMacs systems’.